Osama Alabdi: Never Ignore a Weak Reaction in the Blood Bank
Aug 14, 2026, 11:21

Osama Alabdi: Never Ignore a Weak Reaction in the Blood Bank

Osama Alabdi, Laboratory Specialist at Almoosa Health Group, shared a post on LinkedIn:

“Never ignore even a weak reaction in the Blood Bank.

I recently had an interesting case involving an 85-year-old ICU patient who required a blood transfusion.

I started with routine blood typing and antibody screening using the Bio-Rad Laboratories gel card.

Blood Group:

  • Anti-A: 4 plus
  • Anti-B: 0
  • Anti-D: 4 plus
  • A Cells: 3 plus
  • B Cells: 4 plus
  • Antibody Screen: Negative

My first thought was an ABO discrepancy such as an A₂ subgroup.
Anti-A₁ testing was positive so I looked further.

I performed an antibody identification panel using the tube method.

The results were:

  • 2 plus at Immediate Spin (IS)
  • 3 plus after 15 minutes at room temperature
  • Negative at 37 degrees Celsius and AHG

This pattern suggested the presence of a cold antibody.

A negative antibody screen by gel card does not always rule out a cold antibody.

Every reaction has a reason. Don’t ignore it.”